CALIBRATION TO A MASTER
1.) Insert Allen wrench through holes on
the top of the instrument (in line with eccentrics) and
loosen set screws.
2.) Using the same Allen wrench adjust the two eccentrics
(the eccentric located on the side of the intrument will
adjust the GAUGING COLUMN front to back) (the eccentric
located on the back of the instrument will adjust the
GAUGING COLUMN side to side).
3.) Position the indicator on the unit (refer to
brochure). Take indicator readings first at the bottom of
the master then at the top.
4.) With indicator at the top position adjust the
eccentric to obtain the same reading as when the
indicator is at the the bottom.
5.) Repeat this procedure until readings are the same
from top to bottom of the master. Then lock set screws on
the top of the instrument. It should not be necessary to
readjust again unless the surface plate is not accurate
or the instrument is stored or moved to another location.
Remember
always adjust eccentrics with the indicator in the top
position.
RECEIVING
INSPECTION
1.) Inspect the instument
for any physical damage in shipment. If any notify both
the carrier and manufacturer immediately.
2.) Unpack and remove corrugated spacers from inside the
unit. Special care should be observed to prevent any
excessive dirt, chips, oil or grease from contacting the
gauging column.
3.) (for 18-X model only) The block on at the top of the
unit is a vibration damper and is designed to move freely
on shoulder screws. (DO NOT DISTURB OR REMOVE)
4.) In case there is any deviation from our
specifications, Please contact Us before rejecting or
returning this instrument.
5.) Do not attempt to lubricate the INDI-SQUARE. No
lubrication is necessary or advised under any
circumstances.
6.) Do not attempt to repair, modify or disassemble this
instrument. Any and all repair work must be done at the
factory. Attempts to repair and or disassemble this
instrument may, at the manufactures discretion, void the
one year warranty.
|